Difference between revisions of "826"

From Sensoray Technical Wiki
Jump to: navigation, search
(Calibration errors)
(Calibration errors)
Line 1: Line 1:
 
==Calibration errors==
 
==Calibration errors==
On 826 SDKs earlier than version 3.2.0, analog calibration values will not be applied without J6 (labeled "Calibration Enable") installed. The shunt is intended to protect against accidental overwriting of the calibration values, but in these SDKs it also prevents the reading of those values. If board calibration is incorrect, make sure J6 is installed.
+
On 826 SDKs earlier than version 3.2.0, analog calibration values will not be applied without J6 (labeled "Calibration Enable") installed. A missing shunt is intended to protect against accidental overwriting of calibration values, but in these SDKs it also prevents the reading of those values. This is resolved in SDK version 3.2.0 and above; in these versions the shunt functions as intended and must be installed only when calibrating the board (though leaving it installed all the time is okay).
  
This is resolved in SDK version 3.2.0 and above; in these versions the shunt functions as intended and needs to be installed only when calibrating the board (though leaving it installed all the time is okay). The 826 SDK can be downloaded from the [http://www.sensoray.com/826 826 product page].
+
If board calibration is incorrect, make sure J6 is installed or upgrade to SDK version 3.2.0 or higher. The 826 SDK can be downloaded from the [http://www.sensoray.com/826 826 product page].

Revision as of 09:54, 25 April 2013

Calibration errors

On 826 SDKs earlier than version 3.2.0, analog calibration values will not be applied without J6 (labeled "Calibration Enable") installed. A missing shunt is intended to protect against accidental overwriting of calibration values, but in these SDKs it also prevents the reading of those values. This is resolved in SDK version 3.2.0 and above; in these versions the shunt functions as intended and must be installed only when calibrating the board (though leaving it installed all the time is okay).

If board calibration is incorrect, make sure J6 is installed or upgrade to SDK version 3.2.0 or higher. The 826 SDK can be downloaded from the 826 product page.

Personal tools
Namespaces

Variants
Actions
Toolbox